Ultra Zero Air Generator GT-3000, 3Nl/min

GT Zero Air Generators The GT Ultra Zero Air Generators produce dry and hydrocarbon/CO/NOx/S­Ox free air, using air from oil-free compressors, thus avoiding the need to use traditional bottles that are often complex to change.

Your price
on inquiry
Availability: ask us
Product description

Product description

GT Zero Air Generators

The GT Ultra Zero Air Generators produce dry and hydrocarbon/CO/NOx/S­Ox free air, using air from oil-free compressors, thus avoiding the need to use traditional bottles that are often complex to change. The ergonomic, compact and silent design makes the GT series generators the ideal solution for zero air supply in the following analytical applications:

  • GC with flame ionization detector (FID)
  • GC with nitrogen phosphorus detector (NPD)
  • GC with flame photometric detector (FPD) and pulsed flame photometric detector (PFPD)
  • Total carbon content (TOC) analysis and others

GT Ultra Zero Air generators use five steps to transform ambient air into analytical grade air. The external oil-free compressor (not included) delivers air through a high efficiency coalescent/par­ticulate prefilter that traps any large particles that may damage the system. The prefilter has an automatic purge system, and traps oil, water and any other particles larger than 5 microns in size. Air is next passed into a dual column pressure swing adsorption air dryer to remove water vapor and carbon dioxide from the supply. While one of the PSA columns is active, the other is pressure-regenerating by back-purging with excess air. From the PSA dryer, air is channeled into a scrubber with activated carbon to reduce NOx and SOx levels to less than 1 ppm through an adsorption process. The NOx/SOx scrubbering activated carbon, contained in a cylinder, becomes saturated after one year of service and must be replaced. The air leaving the scrubber enters a high-temperature platinum catalyser, which through oxidation eliminates hydrocarbons and carbon monooxide down to < 0.1 ppm. A final high-grade filter is used to remove 99.99% of particulates greater than 0.5 microns in size.

Better Detector Performance

The reduction of hydrocarbons, including methane, and carbon monoxide to < 0.1 ppm, as well as NOx to < 1 ppm, decreases the background noise level and gives the baseline much better stability, considerably increasing detector sensitivity and ensuring precise analytical results.

Easy Installation

Just connect a suitable oil free compressor to almost immediately have air at the right grade. To save bench space, the unit can be easily installed on the lab wall. No more bottles in the lab and expensive pipelines for air distribution.

Save Money

The unit only requires connection to a suitable oil free compressor and the mains: the investment can be paid back in less than a year.

Technical Specifications
Max. zero air flowrate 3 Nl/min
Max. zero air pressure 6.5 bar
Zero air hydrocarbon content <0.1 ppm
Zero air CO content <0.1 ppm
Zero air CO2 content <5 ppm
Zero air NOx,SOx content <0.1 ppm
Zero air particles removed 99.99% with size >0.5 micron
Zero air dewpoint <-50°C
Zero air temperature Ambient + 15°C
Max. inlet hydrocarbon content 100 ppm
Max. inlet CO content 50 ppm
Max. inlet CO2 content 500 ppm
Max. inlet NOx,SOx content 50 ppm
Max. inlet temperature 40°C
Inlet pressure range 4.5–10 bar
Inlet compressor Not included
Pressure drop <1 bar
Stabilization time after startup 45 min
Inlet/Outlet connection 1/4" NPT / 1/8" NPT
Power 110–120V 60Hz / 220–240V 50Hz
Dimensions (HxWxD) 410×550×250 mm
Weight 9 kg