ENEN CZCZ SKSK PLPL RURU

RESTEK Electronic Leak Detector

on inquiry
List Price
on inquiry
Your price
on inquiry
Catalog number: R*28500

Leak Detector ideal for use to check leakages from GC and GC/MS instruments

– Audible tone indicates the severity of a leak
- Redesigned circuitry offers 12 hours of operation between charges
- Detects a broad range of gases; Ex rated for use with hydrogen and other explosive gases

NEWS
Possible to use while charging
Switch to NiMH battery

Note: not for use in hazardous area|

Leak Detector Specifications

Detectable Gases Helium, nitrogen, argon, carbon dioxide, hydrogen*
Battery Rechargeable nickel-metal hydride (NiMH) internal battery pack (12 hours normal operation)
Ambient Temperature 10 to 37°C
Humidity Range 0 to 97%
Warranty One year
Certifications CE (EU, Korea, Japan, Australia)
Compliance WEEE, CEC, China RoHS 2

Limits of Detection

These gases can be detected with the Restek electronic leak detector at the following leak rates:

Gas Minimum Detectable Gas Limits Indicating LED Color:
Helium 1.0 × 10–5 red LED
Hydrogen* 1.0 × 10–5 red LED
Nitrogen 1.4 × 10–3 yellow LED
Argon 1.0 × 10–4 yellow LED
Carbon dioxide 1.0 × 10–4 yellow LED

Gas detection limits measured in atm cc/sec. *The Restek electronic leak detector should only be used to detect trace amounts of hydrogen in a noncombustible environment. It is NOT designed for determining leaks in a combustible environment.