ENEN CZCZ SKSK PLPL RURU

PC 50 VioLab kit with pH electrode CHS ChemFlex S7, cond.cell CHS CondiLab 22P, standards and access.

on inquiry
List Price
on inquiry
Your price
on inquiry
Catalog number: CHS-5358-C3M

Multimeter PC 50 VioLab Bench multimeter

pH – mV/ORP – COND – Temperature
  • Large backlight display pH/mV/Conductivity and Temperature parameters simultaneously
pH
  • Selectable resolution: 0.1/0.01 pH
  • Automatic calibration up to 3 points with 2 selectable pH-buffer sets (USA and NIST)
  • Manual calibration up to 2 points with user-defined pH-Buffer
  • Stability parameter (Low-Med-High) for high accuracy measurement
Conductivity
  • Automatic calibration from 1 to 5 points between 5 conductivity standard values: 84 µS – 147 uS – 1413 µS – 12.88mS – 111.9 mS
  • Manual calibration at 1 point (User defined)
  • Automatic and manual compensation of temperature with reference value 15…30 °C

PC 50 VioLab DHS complete kit include pH electrode CHS ChemFlex S7, and conductivity cell CHS CondiLab 22P with integrated ATC with 1m cable BNC/Cinch, 1 m S7/BNC cable electrode holder and colored pH buffer solution pH 7 and pH 4, standard calibration solutions 1413 µS and 12,88 mS and 5V/220V AC power supply.

pH
Measuring range 0.00…14.00pH
Resolution 0.1/0.01pH
Calibration points 1 or 3
Buffer recognized USA: 1.68 – 4.01 – 7.00 – 10.01 pH; NIST: 1.68 – 4.01 – 6.86 – 9.18 pH; 2 points defined by user
Indication of the Slope and Offset Yes
Criteria of measurement stability Yes
Indication of calibration point Yes
Potencial measuring range ± 1.000 mV
Potencial resolution 1 mV
ORP
Measuring range ± 1.000 mV
Calibration points 1 (475 mV)
Conductivity
Measuring range 0.0…200 – 2000 µS/20.00–200.0 mS
Resolution 0.1/1 µS to 0.01/0.1 mS
Relative accuracy ±2% full scale
Calibration points 1..5
Calibration standards recognized 84µS, 147 uS, 1413µS, 12.88mS, 111.9mS, 1 point user defined
Cell constant selectable 0.1, 1, 10 cm-1
Temperature coefficient 0.00…10.00 %/°C
Refenece temperature 15…30 °C
Indication of calibration standards Yes
Measuring range 0…200 g/l
  TDS factor/Accuracy 0,40..1,00/±1% full scale
Temperature
Measuring range 0…100.0°C
Resolution/Accuracy 0.1°C / ±0.5°C
Temperature compensation Automatic/manual 0…100°C
Others
Inputs 2 BNC, 2 RCA-CINCH (ATC)
IP protection IP54
Power supply 5V DC/220 V
Dimensions/weight instrument 160×140×45(H)­mm/380g

Use

  • MultiparameterPhys. property
  • StolníInstrument type

Did not find the necessary information? Write to us and we will immediately respond to e-mail.

Laboratory instruments and accessories

Measurement and control

Rozbalit vše

Copyright © CHROMSERVIS s.r.o., Jakobiho 327, CZ-109 00 Praha 10 ~ Tel: +420 274 021 211 ~ Znalostní databáze ~ e-shop od MyWebdesign.cz